Prepare Venous InterGraft

Venous InterGraft delivery system preparation

Follow these steps prior to delivering venous InterGraft connection

  1. Using sterile technique, open VIG package and place InterGraft delivery system in sterile field.
  2. Remove the white graft grippers.
  3. Attach a stopcock to each of the T-B ports D, E.
  4. Loosen the T-B valves A, C.
  5. Using 10cc syringe, flush the delivery system with sterile saline solution. Close valve A
  6. loosen both T-B valves A, C.
    1. Load the 0.018" guide wire into the VIG delivery system.
    2. Backload wire into delivery system and gently close T-B valve C.
  7. Deliver Venous InterGraft